Handheld Flow Analyzer
Handheld Flow Analyzer
Certifier FA PLUS Gas/Mode Range Accuracy**
Flow—High Flow Module Air, O
2
-200 to +300 slpm* ±2% or ±0.075 slpm
Air/O
2
Mixtures 0 to 300 slpm ±4%o0.1slpm
N
2
-200 to +300 slpm ±3%o0.1slpm
CO
2
-40 to +40 slpm ±3%o0.1slpm
Flow—Low Flow Module Air, O
2
0.01 to 20 slpm ±2% or ±0.01 slpm
N
2
,CO
2
0.01 to 20 slpm ±3% or ±0.01 slpm
N
2
O 0.01 to 20 slpm ±4% or ±0.025 slpm
Volume-High Flow-Inhaled Air, O
2
0.01 to 10 liters STP ±2% Plus 0.02 liters
Air/O
2
Mixtures 0.01 to 10 liters STP ±4% Plus 0.02 liters
Volume-High Flow-Exhaled Air, O
2
0.01 to 10 liters STP ±3% Plus 0.03 liters
Air/O
2
Mixtures 0.01 to 10 liters STP ±4% Plus 0.04 liters
Volume-Low Flow-Inhaled Air, O
2
0.01 to 10 liters STP ±2% or ±0.01 liters
N
2
O 0.01 to 10 liters STP ±4% or ±0.01 liters
Minute Volume-High Flow 0.01 to 100 liters STP ±3%
Minute Volume-Low Flow 0to10litersSTP ±3%
Respiratory Times Ti,Tip,Te 0.04 to 30 seconds ±2% or ±0.01 seconds
I:E Ratios I:E, I:Eip 1:100 to 100:1 ±4%
Respiratory Rate f 1 to 1500 breaths per minute ±2% or 0.1 bpm
Low Pressure All -25 to +150 cmH
2
O ±0.5% or ±0.15 cmH
2
O
High Pressure -10 to +150 PSI (-0.7 to 10 bar) ±1% or 0.1 PSI (7 mbar)
Barometric Pressure 7 to 23 PSI (500 to 1600 mbar) ±0.16 PSI (11 mbar)
Oxygen Concentration 21% to 100% 2% of concentration
Specifications
54
54
Flip for product pricing - Page U
Flip for product pricing - Page U
To Order Certifier FA Plus
Model Description
4080 High-Flow Standard Kit
Optional Modules and Accessories
Model Description
4073 Oxygen Sensor Kit
PSR-11-917-J Replacement Oxygen Sensor
4082 Low-Flow Module
Part # Description
1208061 Extra battery pack and charger kit
1303860 Printer cable
1602342 Low Flow Filter
BC20-40702 High Flow Filter
*slpm = Standard Liters per Minute **Accuracy stated as a percent of reading at TSI standard gas conditions.
4080 High-flow test system with
4082 Low-flow kit (sold separately)
Przeglądanie stron 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 63 64 65

Komentarze do niniejszej Instrukcji

Brak uwag