Defibrillator Analyzer
Defibrillator Analyzer
DA-2006
DA-2006
Flip for product pricing - Page E
Flip for product pricing - Page E
Typical DA-CS-06 Setup
Typical DA-CS-06 Setup
Our unique DA-CS-06 Defibrillator Analyzer
Calibration Standard system is available to qualified
customers that have a need for a defibrillator output
energy standard to be used in the calibration of their
analyzers, regardless of manufacturer or model. The
DA-CS-06 was originally developed to be used in the
calibration of our own DA-2006 and DA-2006P ana-
lyzers.
S
S
PECIFICATIONS
PECIFICATIONS
TRANSCUTANEOUS PACEMAKER ANALYZER
DEMAND SENSITIVITY
WAVEFORMS
SELECTIONS
Square
Triangle
Haversine
WIDTH 10,25,40,100,200 ms
ECG OUTPUT
AMPLITUDE – OUT 0 to 4 mv
RESOLUTION – OUT 40 μv
ACCURACY – OUT +/-2%
PACER INPUT (50 TO 400 OHMS)
AMPLITUDE – OUT 0 to 10 mv / 50 Ohms
RESOLUTION – OUT 40 μv
ACCURACY – OUT +/-2%
RATE – IN 30 to 120 ppm
PACER INPUT (500 TO 2300 OHMS & OPEN)
AMPLITUDE – OUT 0 to 100 mv
RESOLUTION – OUT 1mv
ACCURACY – OUT +/-2%
RATE – IN 30 to 120 ppm
DEFIBRILLATOR PLATES
AMPLITUDE – OUT 0 to 10 mv
RESOLUTION – OUT 0.1 mv
ACCURACY – OUT +/-2%
RATE – IN 30 to 120 ppm
TRANSCUTANEOUS PACEMAKER ANALYZER
50/60 HZ INTERFERENCE TEST SIGNAL
ECG OUTPUT
0,0.4,0.8,1.2,1.6,2.0,2.4,2.8,
3.2,3.6,4.0 mv
PACER INPUT > 500 OHMS 0,10,20,30,40,50,60,70,80,90,100 mv
DEFIBRILLATOR PLATES 0,1,2,3,4,5,6,7,8,9,10 mv
TRANSCUTANEOUS PACEMAKER ANALYZER
REFRACTORY PERIOD
PACING 20 to 500 ms
SENSING 20 to 500 ms
ACCURACY +/-2 ms
TRANSCUTANEOUS PACEMAKER ANALYZER
TEST LOAD
RANGE
50,100,150,200,300,400,500,600,700,
800,900,1000,1100,1200,1300,1400,
1500,1600,1700,1800,1900,2000,
2100,2200,2300 Ohms
ACCURACY
50 to 1300 Ohm +/-1%
1400 to 2300 Ohm +/-1.5%
TRANSCUTANEOUS PACEMAKER ANALYZER
OSCILLOSCOPE OUTPUT
0 – 150 V 10.24:1 amplitude attenuation
15 –
60 V 41:1 amplitude attenuation
> 60 V 164:1 amplitude attenuation
MAX OUTPUT 200 V
TRANSCUTANEOUS PACEMAKER ANALYZER
PULSE MEASUREMENTS
AMPLITUDE 4 to 300 mA (100 Ohm load)
ACCURACY +/-5% or +/-0.5 mA
RATE 30 to 800 ppm
ACCURACY +/-1% or 2 ppm
PULSE WIDTH 0.6 to 80 ms
ACCURACY +/-1% or +/-0.3 ms
MAX VOLTAGE
200 V (Variable Load Input Jacks)
15 V (Fixed Load Input Jacks)
ENERGY OUTPUT MEASUREMENT
GENERAL
METHOD Biphasic
LOAD RESISTANCE
50 Ohms +/- 1%,
non-inductive (<1 μH)
DISPLAY RESOLUTION 0.1 Joules
MEASUREMENT TIME
WINDOW
100 ms
ABSOLUTE MAX PEAK
VOLTAGE
6000 Volts
PULSE WIDTH 100 ms
ENERGY OUTPUT MEASUREMENT
HIGH RANGE
VOLTAGE <5000 Volts
MAX CURRENT 120 Amps
MAX ENERGY 1000 Joules
ACCURACY
+/-2% of reading for >100 Joules
+/-2 Joules for <100 Joules
TRIGGER LEVEL 100 Volts
PLAYBACK AMPLITUDE 1 mv / 1000 V Lead 1
TEST PULSE 125 Joules +/- 20%
ENERGY OUTPUT MEASUREMENT
LOW RANGE
VOLTAGE <1000 Volts
MAX CURRENT 24 Amps
MAX ENERGY 50 Joules
ACCURACY
+/-2% of reading for >20 Joules
+/- 0.4 Joules for <20 Joules
TRIGGER LEVEL 20 Volts
PLAYBACK AMPLITUDE 1 mv / 1000 V Lead 1
TEST PULSE 5 Joules +/-20%
ENERGY OUTPUT MEASUREMENT
OTHER
OSCILLOSCOPE OUTPUT
HIGH MEASURE RANGE 1000:1 amplitude-attenuated
LOW MEASURE RANGE 200:1 amplitude-attenuated
WAVEFORM PLAYBACK
OUTPUT LEAD I & PLATES
SCREEN 200:1 Time Base Expansion
SYNC TIME MEASUREMENTS
TIMING WINDOW Starts 40 ms before each R-wave peak
TEST WAVEFORMS All waveform simulations available
DELAY TIME ACCURACY +/- 1 ms
CHARGE TIME MEASUREMENT
From 0.1 to 99.9 sec
ECG GENERAL
LEAD TO LEAD
IMPEDENCE
(RL, LL, RA, LA)
1000 Ohms
LEAD TO LEAD
IMPEDENCE
(V1-V6)
1000 Ohms
ECG ARRHYTHMIA SELECTIONS
Ventricular Fibrillation
Atrial Fibrillation
Second Degree A-V Block
Premature Atrial Contraction
PVC Early
PVC Standard
PVC R on T
Multifocal PVC
Bigeminy
Run of 5 PVCs
Ventricular Tachycardia
CARDIOVERSION
DELAY 0 to 6000 ms
RESOLUTION 0.1 ms
ACCURACY +/-2 ms
ECG NSR
RATE 30 to 300 BPM
ACCURACY +/- 1%
AMPLITUDE 0.5,1.0,1.5,2.0 mv (Lead II)
ACCURACY +/- 2% @ Lead II
HIGH LEVEL 200 times Amplitude
ACCURACY +/- 5%
QRS DURATION 80ms
ECG PERFORMANCE
SINE WAVE 0.1 to 100 Hz
SQUARE WAVE 0.125, 2.000 Hz
TRIANGLE WAVE 2.000, 2.500 Hz
PULSE WAVE 30,60,120 BPM; 60 ms width
AMPLITUDE 0.5,1.0,1.5,2.0 mv (Lead II)
RATE ACCURACY +/- 1%
AMPLITUDE ACCURACY +/- 2% @ Lead II
SHOCK ADVISORY ALGORITHM TEST
ECG SIGNALS
Asystole
Coarse Ventricular Fibrillation
Fine Ventricular Fibrillation
Multifocal Ventricular Tachycardia @ 140 BPM
Multifocal Ventricular Tachycardia @ 160 BPM
Polyfocal Ventricular Tachycardia @ 140 BPM
Polyfocal Ventricular Tachycardia @ 160 BPM
Supra Ventricular Tachycardia @ 90 BPM
DATA INPUT/OUTPUTS
Parallel Printer Port
RS232C (for computer control)
PHYSICAL
DISPLAY LCD Graphical 256 X 64 Pixels,
Backlight
ENCLOSURE
3.4 x 9.8 x 10.7 Inches
(86.4 x 249 x 271.8 mm)
Royalite R59, UL 94V-0
WEIGHT
< 5Lbs
(< 2.3 Kg)
FACEPLATE Lexan, Back printed
OPERATING RANGE 15 to 40 C
STORAGE RANGE -20 to 65 C
ELECTRICAL
POWER
Battery, 9 VDC (2 required)
(NE 1604)
Alkaline
BATTERY ELIMINATOR
(Optional)
BE2006PU (120 VAC) – US
BE2006PE (220 VAC) – Euro
10V, 300 mA DC
BC20-21103 (110 V)
BC20-21101 (220 V)