Anesthetic Agent Analyzer
Anesthetic Agent Analyzer
48
Gas Monitoring
Gas Monitoring
Method: Sidestream; Non-dispersive
infrared (NDIR)
Identified Gases: Halothane,Enflurane, Isoflurane,
Desflurane,
Sevoflurane, CO
2, N2O,
Oxygen
Concentration Units: Vol%, Torr, kPa, mmHg
Flow Rates: 100 ml/min, 150 ml/min, 200
ml/min
Agent Detection
Agent Detection
Measurement Range: Halothane: 0 to 10%
Enflurane: 0 to 10%
Isoflurane: 0 to 10%
Desflurane: 0 to 20%
Sevoflurane: 0 to 10%
CO
2: 0 to 12.5%
N
2O: 0 to 99%
Oxygen: 0 to 100%
Measurement Accuracy: Agents: ± (0.1% abs. + 4% reading
CO
2: ± 0.2% abs. or 4% of reading
N
2O: ± (1.5% abs. + 4% of reading
Oxygen: ± 3 vol% (for 0-90%),
± 4 vol% (for 91-99%)
Time to Detect Agent: < 15 seconds @ 200ml/min
Agent Detection Resolution: 0.1 Volume Percent
Mixed Gas Threshold: 0.2 vol. % +10% of total
concentrations
Rise Time: Agents: 450 msec
CO
2: 350 msec
N
2O: 400 msec
Oxygen: 600 msec
Respiration Rate
Respiration Rate
Range: 1 to 60 breaths/minute
Accuracy: ± 2 breaths/minute or 2% of reading
System Features
Occlusion Clearing: Automatic
Auto Zeroing: Occurs 30 to 60 minutes
Duration: 3.0 to 7.0 seconds
Manual user calibration not
required.
Temperature stabilized optical
assembly
Auto-calibration; verification
recommended
every 6-months.
Warm-up Time: 1 minute to first waveforms;
< 20 minutes to full accuracy
Display
Display
Main Display: Active color TFT
Size: 5.5" Display Area
Resolution: Internal Display; 320 x 240 pixels
External Video; 640 x 480 pixels
Number of Traces: 3 maximum
Controls
Controls
Keys: 7; membrane-activated
Rotary knob: Push and rotate; 24 steps/turn
System Outputs
System Outputs
Com Ports: Com 1; RS232-compatible,
Serial DB-9
Com 2; Serial/Analog, mini-
DIN8
Video Port: Color VGA compatible,
High Density D-Sub 15
Alarms
Alarms
Alarm Characteristics: EN 475, Adjustable; with
audible and visual
indications.
Alarm Levels: High, Medium, Low, Informational
Alarm Modes: Adult/Pediatric/Neonate High
and low limit settings for each
mode.
Trends
Trends
Memory: 24 hours of stored data,
History available with
external VGA display
Display: Tabular, Graphical
Power Requirements
Power Requirements
Voltage: 100, 120, 220, 240 VAC;
50/60 Hz
Power Consumption: 40 W, typical
Battery: NiMH; Life: 1 hour, typical;
Recharge time: 3 hours
Mechanical
Mechanical
Weight: ≤13 Lbs (5.9 kg )
Size: 16.5cm (H) x 27.9cm (W) x
30.5cm (D); 6.5” (H) x 11.0”
(W) x 12.0” (D)
Environmental
Environmental
Operating Temperature: 15 to 35 ºC; 59 to 95 ºF
Storage Temperature: -5 to 50 ºC; 23 to 122 ºF
Operating/Storage Humidity: 15% to 95%, noncondensing
Altitude: -300m to 3000m
(-1,000 to 10,000ft.)
Type of Protection: Class I Equipment
Degree of Protection: Type CF, Defibrillator-proof
Protection against ingress: Ordinary