Patient Simulator Series
Patient Simulator Series
10
10
S
S
PECIFICATIONS
PECIFICATIONS
PS-2210
PS-2210
with Cardiac
with Cardiac
Output Option
Output Option
PS-2220
PS-2220
To Compare Patient Simulators,
see chart on Page 6.
Flip for product pricing - Page C
Flip for product pricing - Page C
ECG NSR
Rate
30,40,45,60,80,90,100,120,140,
160,180,200,220,240,260,280,
300 BPM
Accuracy ±1%
Amplitude
.05,.10,.15,.20,.25,.30,.35,.40,.4
5,.50,1.0,1.5,2.0,2.5,3.0,3.5,4.0,
4.5,5.0 mV
Accuracy ± 2% @ Lead II
High Level 200 times Amplitude
Accuracy ±5%
QRS Interval Adult (80ms), Pediatric (40ms)
ST Segment (Elevation) ± 0,.05,.1,.2,.3,.4,.5,.6,.7,.8 mV
ECG Performance
Sine Wave 0.1,0.5,5,10,40,50,60,100 Hz
Square Wave 0.125, 2.000 Hz
Triangle Wave 2.000, 2.500 Hz
Pulse Wave 30,60,120 BPM; 60 ms
Amplitude
.05,.10,.15,.20,.25,.30,.35,.40,.4
5,.50,1.0,1.5,2.0,2.5,3.0,3.5,4.0,
4.5,5.0 mV
R-Wave (Havertriangle)
Rate 30,60,80,120,200,250 BPM
Width
8,10,12,20,30,40,50,60,70,80,9
0,100,120,130,140,150,160,170
180,190,200 ms
Amplitude
.05,.10,.15,.20,.25,.30,.35,.40,.4
5,.50,1.0,1.5,2.0,2.5,3.0,3.5,4.0,
4.5,5.0 mV
Rate Accuracy ±1%
Amplitude Accuracy ± 2% @ Lead II
R-Wave </= 20 ms ± 5% @ Lead II
Blood Pressure
Channels
1 – PS-2210
2 – PS-2220
4 – PS-2240
Static Pressure
-10,-5,0,20,40,50,60,80,100,
150,160,200,240,250,300,320
400 mmHg
Accuracy ± (2% of Reading + 2 mmHg)
Impedance 300 Ohms
Accuracy ± 10%
Excitation Range 2to16VRMS
Excitation Frequency DC to 5 KHz
Sensitivity 5 or 40 µV/V/mmHg
Respiration Artifact 0 to 16 mmHg
Remote(Option)
Fully compatible with PSR-2200
Pacemaker Waveforms
Amplitude 1,2,3,4,5,6,7,8,9,10 mV
Accuracy ± 10%
Width 0.1,0.5,1.0,1.5,2.0 ms
Accuracy ±5%
Rate 75 BPM
Accuracy ±1%
Respiration
Rate
Apnea,15,20,30,40,60,80,100,1
20 BrPM
Accuracy ±1%
Impedance delta 0.1,0.2,0.5,1.0,2.0,3.0 Ω
Accuracy ± 10%
Baseline 500,1000,1500,2000 Ω
Accuracy ±5%
Lead LA or LL
Temperature
Temperature
0,24,30,37,40 C (32.0,
75.2,86.0,98.6,104.0 °F)
Accuracy ±0.1°C
Type YSI Series 400 and 700
Fetal / Maternal
Fetal Heart Rate
60,90,120,140,150,210,240 BPM
Inter Uterine Pressure
Response Types
Uniform Deceleration
Early Deceleration
Late Deceleration
Uniform Acceleration
IUP Wave
Bell Curve with 90 mmHg
Peak and 90 Sec Width
IUP Trigger
Manual
Auto: 2,3,5 Minute
Cardiac Output
Injectate Volume 10 cc
Injectate Temperature 0°Cor24°C,
±
2%
Calibration Coefficient 0.542 (0 °C Inj), 0.595 (24 °C Inj)
Blood Temperature 37 °C (98.6 °F),
±
2%
Cardiac Output 2.5,5.0,10.0 l/min,
±
5%
Simulations
Normal Flow
Faulty Injectate
Left-to-Right Shunt
Temperature Calibration Pulse
Temperature Calibration
Pulse
1.5 °C Down for 1 sec,
±
1%
Catheter Type Baxter Edwards, 93a-131-7f
SpO
2
Simulation (MSP-2100 & Finger SIM
TM
Req’d.)
Rate
30 to 300 BPM
SpO
2
Output 80, 90 & 97%