RF MEASUREMENT
INPUT
IMPEDANCE
50 Ω
CONNECTION
BNC (50 Ω)
MAXIMUM
VOLTAGE
3.3 V pk-pk
Internally Protected
FREQUENCY
10 kHz 10 MHz
INPUT
COMPATIBILITY
CURRENT
TRANSFORMER
Pearson Electronics (Typical)
ATTENUATION
RATIOS
0.1:1 (Pearson Model 411)
1:1 (Pearson Model 4100)
User Selectable
VOLTAGE (RMS)
100 mV RANGE
0.20 70.00 mV RMS
0.01 mV Resolution
1000 mV RANGE
2.0 700.0 mV RMS
0.1 mV Resolution
VOLTAGE (pk, pk-pk)
100 mV RANGE
0.20 100.0 mV
0.01 mV Resolution
1000 mV RANGE
2.0 1000.0 mV
0.1 mV Resolution
ACCURACY,
100 mV RANGE
f ≤ 1.0 MHz
1.0 MHz < f
2.5 MHz
f > 2.5 MHz
Input
≤ 10.0 mV
± 1% Reading
Or ± 0.25 mV
± 1% Reading
Or ± 0.25 mV
± 1% Reading
Or ± 1.0 mV
10.0 mV < Input ≤
35.0 mV
± 1% Reading
Or ± 0.25 mV
± (1% Reading
+ 0.5% Range)
± (6% Reading +
0.5% Range)
Input
> 35.0 mV
± 1% Reading
Or ± 0.25 mV
± (1% Reading
+ 0.5% Range)
± (6% Reading +
5% Range)
ACCURACY,
1000 mV RANGE
ƒ ≤ 2.5 MHz
ƒ > 2.5 MHz
Input ≤ 50 mV
± 1% Reading
Or ± 1 mV
± 1% Reading
Or ± 1 mV
50 mV < Input ≤
400.0 mV
± 1% Reading
± 4% Reading
Input > 400.0 mV
± (1% Reading +
0.25% Range)
± (4% Reading +
0.25% Range)
SPECIFICATIONS