CONFigure Subsystem
This group allows the user to setup the display and operational settings for the unit.
KEYWORD PARAMETER FORM COMMENTS
CONFigure
:DISPlay :SxZy S<display_screen_number> Z<zone_number> nn
display_screen_number = 1-7
nn = Parameter for selected Zone
6 = Measurement List Display (Non-editable)
7 = Quick Configuration Screen (Non-editable)
:SCReen < numeric_value >
:AVERaging FAST | SLOW | MEDium
:HOLD ON | OFF
:LOAD
:MODE TABle | ADJustable
:VALue < numeric_value > Table Mode: 1-12 for individual
resistors, 13-22 for resistor
sets 1-10
Adjustable Mode: 0-6500.0
:SETn < numeric_value > n = Resistor set to configure
(1-10)
< numeric_value > = 16 bit
binary value of resistors to
include in set ‘n’ where bit
0 = Load 1, Bit 1 = load 2…
Bit 11 = Load 12
each bit selects whether the
load is included in the set.
Bit = 1 includes the load
Bit = 0 excludes the load.
:NUMber < numeric_value > set of valid numeric values
:Ln < numeric_value > n = Load to configure
< numeric_value > = Actual
value of load ‘n’;
0-6500.0 ohms
INPut:
Atten: 0.1 | 1
RANGe: 100 | 1000 | AUTo
NUMsamples: 1024 | 2048 | 4096 | 8192 | 16384 | 54768
MODE: CONTinous | PULs
6=Measurement List Display