The following is a breakdown of the parameters available in the LOAD SETUP MODE
and their available options:
Load Setup Configuration
Parameter
Description
Range
Number Of Loads
Sets the number of load resistors present in the
system. This determines the maximum
combination of resistors available when the Load
Selection is set to Table.
1-12
Load 1
Calibrates the Load 1 Resistance Value. This
should be set to the actual resistance of the
smallest resistor in the system.
0.0-6500.0 ohms
Load 2
Calibrates the Load 2 Resistance Value. This
should be set to the actual resistance of the next
higher resistor in the system.
0.0-6500.0 ohms
Load 3
Calibrates the Load 3 Resistance Value. This
should be set to the actual resistance of the next
higher resistor in the system.
0.0-6500.0 ohms
Load 4
Calibrates the Load 4 Resistance Value. This
should be set to the actual resistance of the next
higher resistor in the system.
0.0-6500.0 ohms
Load 5
Calibrates the Load 5 Resistance Value. This
should be set to the actual resistance of the next
higher resistor in the system.
0.0-6500.0 ohms
Load 6
Calibrates the Load 6 Resistance Value. This
should be set to the actual resistance of the next
higher resistor in the system.
0.0-6500.0 ohms
Load 7
Calibrates the Load 7 Resistance Value. This
should be set to the actual resistance of the next
higher resistor in the system.
0.0-6500.0 ohms
Load 8
Calibrates the Load 8 Resistance Value. This
should be set to the actual resistance of the next
higher resistor in the system.
0.0-6500.0 ohms
Load 9
Calibrates the Load 9 Resistance Value. This
should be set to the actual resistance of the next
higher resistor in the system.
0.0-6500.0 ohms
Load 10
Calibrates the Load 10 Resistance Value. This
should be set to the actual resistance of the next
higher resistor in the system.
0.0-6500.0 ohms
Load 11
Calibrates the Load 11 Resistance Value. This
should be set to the actual resistance of the next
higher resistor in the system.
0.0-6500.0 ohms
Load 12
Calibrates the Load 12 Resistance Value. This
should be set to the actual resistance of the next
higher resistor in the system.
0.0-6500.0 ohms