GETTING READY FOR A FLOW--PCA TEST
GETTING READY FOR A FLOW--PCA TEST
INFUSION INFUSION
INFUSION INFUSION
INFUSION
ANAN
ANAN
AN
ALAL
ALAL
AL
YZER--FLYZER--FL
YZER--FLYZER--FL
YZER--FL
OO
OO
O
WW
WW
W
Choose Channel Screen
To run a test, toggle to the appropriate Channel screen
using .
STEP 2
9
Connect and Prime Device
Connect device under test (DUT) to the unit and prime the
channel. (See PRIMING for details.)
There is an optional PCA output for connecting the module
to the PCA input of the DUT. The module will simulate patient
request when there is no flow.
STEP 1
9
NOTES:
C
The NORMAL LED on the remote
module will blink at a rate of twice per
second, when a channel needs to be
primed.
C
As the channel becomes ready, the
NORMAL LED on the remote module
will blink at a rate of once per second.
C
The priming message will not display
if the DUT is already primed.
NOTES:
C
There can be up to two modules
connected to the analyzer.
C
The modules are interchangeable and
the unit will run with a single module
connected to either port.
C
The Channel 1 or 2 screens will only
be displayed when the modules are
connected to those ports.
C
These instructions will utilize
Channel 1 for examples; Channel 2
functions identically.
CHANNEL 1 IDLE
PRIME WITH 50 mL
THEN PRESS START