PURGING OVERVIEW
PURGING OVERVIEW
PURGINGPURGING
PURGINGPURGING
PURGING
This section covers how to Purge the unit after use. Only
Distilled Water is to be used in the system; this should keep it
clean. However, if no fluid tests are going to be run for a
period of more than 24 hours, the water must be blown out of
the unit to maintain the necessary cleanliness.
PURGING
CAttach Clean Dry Air (CDA) or N
2
Line to Fluid In on Module
CAllow CDA or N
2
to Flow at 20 PSI for At Least One Minute
CMove Line to Fluid Out on Module
CGo Into User Setup Mode
CSet Air Purge Parameter to On
CWait at Least 2 Minutes Before Removing Line
The following is an outline of the steps that are involved in
purging the unit: