STEP 3
9
Stop the Test
Press when the test is completed.
The test is also terminated by a NURSE CALL.
STEP 4
9
View Datalog
When the test is completed, the screen will automatically
change to display the Datalog for that test.
(See DATALOG--OCCLUSION TEST for details.)
CHANNEL 1 DIST OCCL
OCCL PRESS 0.0PSI
MAX PRESS 0.0PSI
ELAPSED TIME 00:00:00
INFUSION INFUSION
INFUSION INFUSION
INFUSION
ANAN
ANAN
AN
ALAL
ALAL
AL
YZER--OCCLYZER--OCCL
YZER--OCCLYZER--OCCL
YZER--OCCL
USIONUSION
USIONUSION
USION
NOTE:
C
Three readings are taken after a
NURSE CALL message. This is done
to confirm that the DUT shut down after
the call.
If a printer is connected, use to get the printout of
the current test’s Datalog.
NOTES:
C
A NURSE CALL reading is time
stamped and recorded in the Datalog.
CDatalogs are stored and can be
printed out at any time.
CIf Autoprint is selected, the printout
will automatically print when the test
is completed.
(See USER SETUP for details.)
RUNNING A DISTAL OCCLUSION TEST