MODEL SELECTIONMODEL SELECTION
MODEL SELECTIONMODEL SELECTION
MODEL SELECTION
MODEL SELECTION
Go into Model Select Mode
When the cursor is under the desired information line,
use to enter the Select mode.
STEP 5
9
NOTE:
CIn the Select mode, the flashing line
cursor will change to a box.
CHAN 1 MODEL - SELECT
MANUF: ABC
MODEL: 123A
S/N: 12345
Scroll to Desired Choice for Information Line
Use to scroll to the desired identification
information. If found, continue to Step 7. If the desired
information is not available, go to the Model Edit mode to
enter it (See Steps 6a through 6c).
STEP 6
9
NOTES:
CThere are 10 Manufacturers stored
at any given time.
CEach Manufacturer has 10 Models
and Serial Numbers associated
with it.
CAs the Manufacturers are scrolled
through, the Models and S/Ns will
adjust as well.
CThe the Model and S/N can be
selected by repeating the steps
used to set the Manufacturer.
CHAN 1 MODEL - SELECT
MANUF: DEF
MODEL: 123B
S/N: 10000
The unit has 100 data sets available for the storage of model
information. They are configured with 10 manufacturer
positions with 10 model / serial number sets for each.
NOTE:
CDuplicate manufactures can be
entered, thereby allowing for many
models or serial numbers from one
manufacturer to be stored.