3
recommended to use an ambient testing temperature of 24.0 ±3.0 °C (75.2 ±5.4 °F).
Sonic energy agitates the water surface through heating and scattering. The time
duration of each test should be limited to a few minutes. Prolonged testing, particularly
at higher power levels, will show visible signs of air bubbles on the transducer, target,
and the absorbing rubber surface.
Transducer Wetting and Placement
To avoid introducing air into the degassed water, insert the transducer at a 45° angle,
then position so it is facing the target. Verify that the transducer surface is uniformly
wetted, if not, wipe the surface clean using your finger. The transducer should be
pointed toward and center-positioned directly above the cone target. Small
measurement variations will occur due to placement of the transducer. Try various
positions above the target to minimize the magnitude of error due to positioning.