Verification of Proper Scale Functioning
Small variations of water surface motion, air currents, or mechanical movements may
cause uncertainties in power measurements. To test scale accuracy at low levels, set
up the scale as in the Operating Procedure section. Place the 1 gram weight on the flat
surface of the target arm. Read meter three times; readings should be within ± 0.2
counts (for example, 14.4 to 14.8). In case of doubt about lower power resolution,
repeat the same procedure using light objects such as thin paper slices to produce
readings of 5 to 10 counts; repeat readings. Average uncertainty should be within ± 0.2
counts on the Watts scale. Avoid mechanical and air movement or variations in
magnetic fields while making tests.
Out of Measurement Range Warnings
Model USP-100A accommodates weight up to 410 grams. When the scale exceeds
this range, “Error 8.3” will be displayed. Something may be pressing against the target
or support. “Error 8.4” indicates underweight condition. If no obvious error has been
made by the user the unit should be returned for service when any code is displayed.
No Display
1. Make sure the AC adapter’s plug is fully seated in the jack at the back of the unit.
2. Use a voltmeter to verify the adapter is producing approximately 12 VDC.
3. Call our service department for assistance.