The Ultrasound Wattmeter, Model USP-100A, is designed to measure the ultrasound
power output of diagnostic or therapeutic transducers up to 30 Watts. The principle of
measurement is the radiant force method. The USP-100A uses a positioning clamp to
hold the transducer in degassed water above a conical target. The ultrasonic energy
passes through the water to reflect off the target and is then absorbed by the rubber
lining. The radiant power is directly proportional to the total downward force (weight) on
the target. This force is then transferred through the target support assembly to the
electro-mechanical load cell inside the scale. The cell is in a computer-controlled
feedback loop and produces a digital readout; Watts of power (custom units) or grams
of force. The choice of units (Watts or grams) is selected by front panel pushbuttons.
INTRODUCTION