External Point to Point Measurements
The unit has the ability to measure Leakage Current and Resistance between two
points, utilizing two test leads. These tests are separated because they use a slightly
different setup than the previous tests.
Point to Point Leakage Current
With the External function selected, the display will show the leakage current between
the test leads.
The test requires that the Chassis Test lead be plugged into the Chassis Connector and
the External Test lead be plugged into the External Connector. The other ends of the
leads are then attached to the points of interest.
Any current flowing between the test points is passed through a 1000 load with either
AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load Selection
switch.
Point to Point Resistance
With the Earth Resistance function selected, the display will show the resistance
between the two test leads.
The test requires two Chassis Test leads, one in the Chassis Connector and the second
in the External Connector. The opposite cable ends should be connected to the points
of interest.
NOTE: Remove any device plugged into the Analyzer Test Receptacle. If there is a DC
voltage present between the two test points, the reading may contain an error. This can
be checked by reversing the connections. If the readings differ, average the two to get
the actual resistance value.