Display
The main information in the system is presented in the 3-1/2 digit LCD display. This
data is provided as simple meter readings with the measurement units indicated to the
right by one of four LED indicators.
Function Selection
Six LEDs and two keys make up the Function Selection Section. The keys are up and
down arrows. When depressed, they step the Analyzer through the available options.
The LED next to the currently selected option is illuminated.
Load Selection
The unit may either use the AAMI ES1-1993 or IEC 601 Test load for measurements.
This is selected by the Load Selection switch.
Analyzer Test Receptacle Control
There are four keys and 8 LEDs in the Analyzer Test Receptacle Control Section. They
allow the manual control of the power connections that are made to the DUT. Internally,
a series of relays are switched by the microprocessor based on the keys that are
depressed. The LEDs indicate the current state of the power connections to the
Receptacle.
Note: The Forward/Reverse Polarity key has a safety delay feature, preventing damage
to the internal relays and the DUT. When the key is depressed, the DUT power is
disabled and the safety delay is activated. When this delay is complete, the internal
relays switch the polarity to the DUT and apply power. This delay allows any reactive
power stored in the DUT to self-discharge before the polarity is reversed.
Note: The unit will power up with the Neutral and Ground Closed, in Forward Polarity
and with the Hot Open. It is recommended that the unit be returned to this condition
when plugging and unplugging the DUT.
Power Outlet Indicators
Three Neon indicators help verify the polarity and wiring of the wall receptacle that the
Safety Analyzer is plugged into.
Self-Test Snaps
There are two snaps on the side of the unit that allow for a quick self-test of the
Analyzer. They provide a fixed 1.0 Ohm resistance to Earth/Ground and a 100 µAmp
leakage current source to Earth/Ground when the Chassis lead is applied and the
Analyzer set to the proper mode.