2-41
Alarms that are silenced but still present can be viewed from
the Main Screen. Press to scroll through alarms from
the Main Screen. The first line of the alarm displays the device
for which the alarm sounded (Channel 1, Channel 2, Safety
Analyzer, System). The second line displays the specific
alarm message.
IPA-2000S
CHANNEL 1
STOPPED-BUBBLE FOUND
ALARMSALARMS
ALARMSALARMS
ALARMS
ALARM OVERVIEW
A special feature is incorporated to automatically scroll through
active alarms in the Main Screen. The ALARM SCAN
parameter in the USER SETUP determines the rate at which
active alarms are automatically scrolled in the Main Screen.
(See USER SETUP for details.)
BUBBLE RECOVERY NOTE:
CWhen bubbles are detected in the system, an automatic bubble recovery routine
will attempt to clear them. The display will indicate that the system is in the
Bubble Recovery Mode. Allow the fluid to continue to flow during this process.
When the bubbles are cleared, it is advisable to run a short test to insure proper
operation.
CIn most cases, it is better to simply Purge and Prime again (See PURGING
and PRIMING for details), being certain that no bubbles get into the unit. This
will insure maximum accuracy and reliability.