ii
WARNING - USERS
The DPM-2000 Series is for use by skilled
technical personnel only.
WARNING - USE
The DPM-2000 Series is intended for testing only
and should never be used in diagnostics,
treatment or any other capacity where it would
come in contact with a patient.
WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the DUT to the DPM-2000
Series. A serious hazard may occur if the patient
is connected when testing with the DPM-2000
Series. Do not connect anything from the patient
directly to the DPM-2000 Series or DUT.
WARNING - MODIFICATIONS
The DPM-2000 Series is intended for use within
the published specifications. Any application
beyond these specifications or any unauthorized
user modifications may result in hazards or
improper operation.
CAUTION - INSPECTION
The DPM-2000 Series should be inspected before
each use for wear and should be serviced if any
parts are in question.
CAUTION – MEDIA COMPATIBILITY
The DPM-2000 Series is intended to be used with
only non-corrosive, non-ionic, or otherwise pure
fluids and/or gases that are compatible with
sensor materials including glass, silicon,
ceramic, epoxy, RTV, gold, aluminum and nickel.